Description
PRODUCT DESCRIPTION
RKk500-12 PH sensor is a good solution to measure the PH value. It does not require a professional calibration instrumentand is simple to operate. lt can be used to continuously measure the PH value of wastewater . It is suitable for agriculture,sewage treatment plants, chemical industry, printing and dyeing, papermaking, pharmaceuticals. , electroplating and environmental protection.
SPECIFICATIONS
Item |
Technical Specification |
Measurement Principle |
Electrochemistry |
Range |
0-14PH |
Supply |
7-30VDC(power consumption<0.2W) |
Accuracy |
±0.05PH |
Resolution |
0.01PH |
Temperature Compensation |
Temperature compensation is optional |
Response time |
<8s(flowing liquid), <14s(stationary liquid) |
Stability |
≤0.01PH/24h |
Output Signal |
4-20mAÂ & RS485Â at the same time |
Calibration Cycle |
Every 6 month(General water ) |
Operating Environment |
0-+80℃(<0.6MPa), high pressure is customizable |
Cable length |
5m(default),customizable |
Ingress Protection |
IP68 |
Storage |
10-60℃@20%-90%RH |
PH SCALE | |||
PH value |
Description |
PH value |
Description |
<4.5 |
Strongly acidity |
7.5-8.5 |
Faintly alkalinity |
4.5-5.5 |
Acidity |
8.5-9.5 |
Alkalinity |
5.5-6.5 |
Faintly acidity |
>9.5 |
Strongly alkalinity |
6.5-7.5 |
Neutral |
PH electrode is not used at ordinary times and can be soaked in 3mol/l KCL solution or saturated KCL solution. It is strictly prohibited to immerse the electrode in distilled water and deionize the water or tap water with minimal plasma content.If the PH electrode is contaminated with inorganic substances, it can be cleaned with 0.1mol/l Hcl or NaOH solution for a few minutes and then washed with distilled water. If the PH electrode is contaminated with organic substances, it can be cleaned with alcohol or acetone and then cleaned with distilled water.(note: the protective cap before the electrode should be removed when using);
Mounting bracket(length=1m) is optional
Reviews
There are no reviews yet.